Working Mode | Rate/VTBI, Time/VTBI, Rate/Time, Micro, Simple Rate, Programmable, DOSE Weight, Drug Lab, AutoBolus, Manual Bolus. | |
Syringe | Open System for 5ml,10ml,20ml,30ml,50ml/60ml 5 kinds of syringe of different brand | |
syringe management | could add / remove / edit / calibrate more than 50 kinds brand syringes | |
Accuracy | ≤ ±2%,Built-in:≤ ±1% | |
Calibrate | Automatic or manual calibration of syringe accuracy | |
Rate | 5ml syringe: 0.01-150 ml/h 10ml syringe: 0.01-300 ml/h 20ml syringe: 0.01-400 ml/h 30ml syringe: 0.01-600 ml/h 50ml (60 ml compatible) syringe: 0.01-1200ml/h |
|
VTBI | 0.1~9999 ml or OFF | |
Total Volume Σ | 0-9999.9ml | |
KVO | 0.1~20ml/h Adjustable,Default:1ml/h | |
Programmable | Maximum 8 steps of different syringe rate / time / VTBI | |
BOLUS | Manual Bolus:long Press the Bolus key to infuse and release to stop Auto Bolus:short press the Bolus key and set the VTBI & Rate | |
Pressure | 200mmHg - 1200mmHg, 20 level, live pressure display, mmHg & kPa two units | |
Pre-Alarm Occlusion | When the pressure in the tube goes up, it has the function of early warning | |
Back off | The pump will automatic release the pressure of syringe tube after occlusion in case the medical risk. | |
Pressure fast drop | Monitoring and controlling safety for the needle loose caused the pressure of syringe rapidly drop. | |
Pre-Alarm Finish | 1 min~30 min, The alarm can be set ahead of VTBI | |
Fast start | First time install the syringe and when the pump start to working, it will automatic remove the mechanical gap for high accuracy. | |
Parameters Modify | changes the Rate, Time, Dose, VTBI & other important parameters during the pump working | |
Keypad Locker | Manual / Auto lock keyboard | |
History Record | Record 20000 infuse records, 10000 alarm records, can be downloaded to the computer to save &query | |
Self-Test | safety live time self-test function to ensure the pump working properly | |
Alarm | Occlusion, Finish, Near Finish, Pause timeout, Pusher error, Lower Battery or damage, Motor error, Incorrect syringe type, handle clip error, parameters error and etc. 20 kinds safety monitoring alarm | |
Light Tips | 3 color light tips:Green for pump working;Orange for low level alarm (counting working); Red for high level alarm (pump stop). | |
Voice Tips | Compliance YY 0709-2009 medical standard, Alarm and key volume level adjustable, with Mute function, | |
Monitor | High brightness color LCD, Adjustable brightness | |
Communication | RS485 port, connect with PC to update / download | |
AC Power | AC100V~240V,50Hz/60Hz,25VA | |
Battery | Ni-MH battery, DC9.6V, 2000mAh (after full charged, could continue work more than 7 hours by 5ml/h rate) | |
Battery power query | Live-time query battery life and battery charge status | |
Class | Class I, Type CF, IPX4 | |
Installation | For round and square pole, Horizontal / Vertical, could superposed 1-3 units | |
Dimension | 1.9Kg , 140mm×145mm×325mm | |
Environment | 5℃~40℃; 20%~90% humidity; 70kPa~106kpa atmospheric | |
Use Range | For the patients with constant intravenous infusion of drugs |